Inductively coupled plasma mass spectrometry for elemental impurities (ICH Q3D, USP <232>/<233>)
ICP-MS (elemental impurities)
ICP-MS digests a sample in acid, sprays it into an argon plasma hot enough to break every molecule into charged atoms, and counts the ions of each element by mass, which lets a laboratory measure trace metals such as lead, cadmium, arsenic and mercury against the limits set by ICH Q3D.
Not offered by Tested.ph
What it answers
- How much of each targeted element is in the sample, usually in micrograms per gram (ppm) or nanograms per gram (ppb)
- Whether the four Class 1 elements (arsenic, cadmium, mercury, lead) are below the concentration a laboratory has derived from ICH Q3D limits
- Whether metal catalysts or reagents used in manufacture, such as palladium, remain at measurable levels
What it cannot answer
- Identity, purity or amount of the peptide
- The chemical form of an element (for example, which arsenic compound), unless a separation step is added
- Elements that were not on the laboratory's target list
- Endotoxin, sterility or residual solvents
What it outputs
- Ion counts for selected isotopes of each target element
- Calibration curves from multi-element standards
- A results table in µg/g or ng/g, with spike recoveries and limits of quantitation
- Instrument checks such as drift between calibration standards run before and after samples
On a COA
Typically a table of elements with a result and a limit, for example "Pb" with a value in µg/g (ppm) and "Complies", or a single line "Elemental impurities (ICH Q3D): complies". Older certificates may show "Heavy metals" with one combined figure, which is a different and less specific test.
What does ICP-MS do to a sample?
The sample is first digested, usually in strong acid in a closed vessel heated by microwaves, so every element ends up dissolved. The solution is sprayed into an argon plasma, which breaks molecules apart and strips electrons from the atoms, and the positive ions pass into a mass spectrometer that counts them at each element's characteristic isotope masses. Counts are converted into concentrations against calibration standards.
A 2024 validation study for elemental impurities in a tablet product used this sequence, noting that microwave digestion with ICP-MS is widely used because it is more sensitive than ICP optical emission spectrometry or atomic absorption, and choosing digestion conditions that reduce the loss of volatile mercury (Hao and colleagues, Journal of Analytical Methods in Chemistry, 2024).
Which elements matter, and who sets the limits?
ICH Q3D, the international guideline on elemental impurities, sorts elements into classes by toxicity and likelihood of being present. Class 1 is arsenic, cadmium, mercury and lead, which the guideline calls human toxicants with limited or no use in manufacture, usually entering through materials such as mined excipients. Class 2A is cobalt, nickel and vanadium. Class 2B covers elements such as palladium, platinum and silver, which may be left out of a risk assessment unless intentionally added. Class 3 includes copper, chromium, tin and others with low oral toxicity (ICH Q3D(R2), 2022).
ICH Q3D sets limits as permitted daily exposures (PDEs) in micrograms per day, with separate values for oral, parenteral and inhalation routes. The table shows the Class 1 values from Appendix 2 of the guideline.
| Element | Oral PDE | Parenteral PDE | Inhalation PDE |
|---|---|---|---|
| Cadmium (Cd) | 5 | 2 | 3 |
| Lead (Pb) | 5 | 5 | 5 |
| Arsenic (As) | 15 | 15 | 2 |
| Mercury (Hg) | 30 | 3 | 1 |
A PDE is a quantity per day, not a concentration. ICH Q3D converts it into a concentration limit by dividing by the daily intake of the product (its simplest option assumes 10 grams a day), and defines a control threshold at 30% of the PDE, below which additional controls are not required. The same PDE can therefore give different µg/g limits for different products, so a certificate should state the concentration limit it used.
How do USP <232> and <233> relate to ICH Q3D?
In the United States, USP General Chapter <232> sets elemental impurity limits and <233> sets procedures. A 2020 study of elemental impurities in drug samples notes that a version of <232> harmonised with ICH Q3D became official in December 2017, that <232> specifies 24 elemental impurities, and that <233> describes determination by either ICP optical emission spectrometry or ICP-MS (Heliyon, 2020). ICH Q3D itself says pharmacopoeial procedures or suitable alternatives should be used, and that unless otherwise justified the test should be specific for each element identified for control.
Why was the old heavy metals test replaced?
The traditional heavy metals test was a colour reaction that gave one combined result for several metals without saying which were present. A 2021 study of elemental impurities in calcium carbonate notes that ICH Q3D moved control from the traditional heavy metals test to a risk-based assessment using modern instruments, and that colorimetry is not adequate for controlling low levels of elemental impurities (Frontiers in Chemistry, 2021). ICP-OES and ICP-MS both measure each element separately.
Are elemental impurities a concern for synthetic peptides?
Synthetic peptides are in scope. The European Medicines Agency's draft guideline on synthetic peptides says synthetic peptides are included in the scope of ICH Q3D, lists elemental impurities among typical specification tests (for example, where a metal catalyst is used), and classes them with process reagents and residual solvents as non-peptide impurities (EMA draft guideline on synthetic peptides, 2023). In regulated manufacture the starting point is a risk assessment rather than routine testing of every batch for every element.
Does Tested.ph offer heavy metals testing?
No, and it is not planned. The standard package covers identity, purity and amount, with endotoxin as an option, all performed by the partner laboratory when testing opens. See what a peptide test can and cannot tell you for the full list of what a record covers.
Sources
- ICH Q3D(R2) Guideline for Elemental Impurities, International Council for Harmonisation
- Development and validation of analytical procedure for elemental impurities in rosuvastatin calcium tablets by ICP-MS and microwave digestion (Hao and colleagues, 2024), Journal of Analytical Methods in Chemistry, via PubMed Central
- Elemental impurities analysis in name-brand and generic omeprazole drug samples (2020), Heliyon, via PubMed Central
- Elemental impurities in pediatric calcium carbonate preparations: high throughput quantification and risk assessment (2021), Frontiers in Chemistry, via PubMed Central
- Guideline on the Development and Manufacture of Synthetic Peptides (draft, EMA/CHMP/CVMP/QWP/387541/2023), European Medicines Agency
Last reviewed 6 Oct 2026.